EasyManua.ls Logo

Philips BiPAP AVAPS - 6.1.9 Replacing the Blower Cap; Figure 6-13: Blower Cap; 1. Remove the UI Knob. Refer to section 6.1.2.; 2. Remove the Top Cover. Refer to Replacing the Top Cover.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 6-14
1118824, VER. 00
6.1.9 REPLACING THE BLOWER CAP
FIGURE 6-13: BLOWER CAP
To remove the Blower Cap:
1. Remove the UI Knob. Refer to section 6.1.2.
2. Remove the Top Cover. Refer to Replacing the Top Cover.
3. Remove the Main PCA. Refer to Replacing the Main PCA.
Included in Kit Tools Required
Blower Cap
#4 x 1/2” screw (x7)
T8 Torx screwdriver
T15 Torx screwdriver
Torque screws to 7 in-lbs
during installation.