Philips BiPAP AVAPS - 6.6 System Run-in; 1. Connect the 0.25” Test Orifice to the Outlet Coupler of the device.; 2. Enter the Provider Setup Mode and set the device’s output pressure to its maximum setting.; 3. Turn on the device’s Blower and allow it to run-in for a minimum of eighteen (18) minutes.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 6-291118824, VER. 00
6.6 SYSTEM RUN-IN
1. Connect the 0.25” Test Orifice to the Outlet Coupler of the device.
2. Enter the Provider Setup Mode and set the device’s output pressure to its maximum setting.
3. Turn on the device’s Blower and allow it to run-in for a minimum of eighteen (18) minutes.