Philips BiPAP AVAPS - 7.2.5 Replacing the Flip Lid Assembly; Figure 7-8: Flip Lid Assembly; 1. Remove the Humidifier Tank Assembly. Refer to Replacing the Water Chamber Assembly.; 2. Using a screwdriver or similar probing tool, push in on the latches that secure the Flip Lid Assembly to the Humidifier Top Housing. Refer to Figure 7-9.

Philips BiPAP AVAPS
126 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PAGE 7-91118824, VER. 00
7.2.5 REPLACING THE FLIP LID ASSEMBLY
FIGURE 7-8: FLIP LID ASSEMBLY
TO REMOVE THE FLIP LID ASSEMBLY :
1. Remove the Humidifier Tank Assembly. Refer to Replacing the Water Chamber Assembly.
2. Using a screwdriver or similar probing tool, push in on the latches that secure the Flip Lid Assem-
bly to the Humidifier Top Housing. Refer to Figure 7-9.
FIGURE 7-9: FLIP LID ASSEMBLY REMOVAL
Included in Kit Tools Required
Flip Lid Assembly
Tank Top Seal
None