PAGE 6-71118824, VER. 00
6.1.3 REPLACING THE RIGHT SIDE COVER
FIGURE 6-5: RIGHT SIDE COVER
TO REMOVE THE RIGHT SIDE COVER
1. Push the locking tab on the end of the Right Side Cover towards the device’s Outlet Port.
2. Lift the Cover away from the device.
FIGURE 6-6: RIGHT SIDE COVER LOCKING TAB
TO REMOVE THE RIGHT SIDE COVER
1. Insert the Right Side Cover, support tabs at the front of the device first, into it’s mounting location.
2. Press the Right Side cover fully into place. Verify that the locking tab snaps and secures the cover.
Included in Kit Tools Required
Right Side Cover None
NOTE
The Right Side Cover is used when there is no
Humidifier present.