2 Understanding Configuration
11
For a complete list of Hardware Settings, see “Hardware Settings” on page 95, or refer to the Service
Guide of your monitor model provided on the Documentation DVD supplied with your monitor.
Entering and Leaving Configuration Mode
Switching between Monitoring Mode and Configuration Mode does not affect the Active Settings.
You can even continue to monitor patients while in Configuration Mode. The password for
Configuration Mode is given in the monitor’s service documentation.
To enter Configuration Mode:
1 In the Main Setup menu, select Operating Modes.
2 Select Config and enter the password.
The monitor displays
Config in the center of the Screen while you are in Configuration Mode.
Before you leave the Configuration Mode, always be sure to save any changes you made under
Main
Setup
> Defaults > Save Defaults.
To return to
• the previously saved User Defaults, select
Load Defaults.
• the Factory Defaults, select
Factory Defaults.
WARNING
If you are handing over the monitor to the end-users directly after configuration, make sure that it is in
Monitoring Mode.
To leave Configuration Mode either:
1 In the Main Setup menu, select Operating Modes, and then select the operating mode you require, or
2 select the gray Config block in the center of the Screen or
3 switch the monitor off, then switch it on again.
• The monitor performs a Hot Start in Monitoring Mode with the same settings (Active Settings), if
it is switched off and on again within one minute.
• The User Default Settings are loaded if the monitor is switched on again after more than one
minute, and the
Automat. Default setting is set to Yes, see “Automat. Default” on page 86.