6 Monitor Settings
71
Configuring NST Report
Monitor Settings: Main Setup > NST Report > Setup
NST Report Factory Default Settings
1
only available if Baseline Limits is set to Manual
2
only available if Variability Lim is set to Manual
NST Report Configuration Implications
Only available on monitors with option #C71 (NST Trace Interpretation).
NST Analysis
Use the NST Analysis setting to switch the NST Report trace interpretation On and Off.
Choices are
On and Off.
NOTE
•When NST Analysis is switched On, the NST Timer will also be switched On.
•When the
NST Timer is switched Off, NST Analysis will also be switched Off.
Guideline
The Guideline setting defines the NST Report guideline. The only choice so far is NICHD v01.
NOTE
The Guideline setting is actually a Global Setting. A setting change does not have to be changed if you
leave the Configuration Mode. It takes effect immediately.
Setting Item Mode Factory Default Choice/Range
NST Analysis C, M On On, Off
Guideline C NICHD 2008, v01 NICHD 2008, v01
Report Recording C, M Immediately Manual, After Recorder Stop, Immediately
Auto Trace Rec. C No Yes, No
Short Term Var. C Off On, Off
Max. Decelerat. C 00 to 15
Min. Accelerat. C 20 to 20
Baseline Limits C Guideline Manual, FHR Alarm Limits, Guideline
Baseline High
1
C 150 bpm 120 bpm to 210 bpm, in steps of 10 bpm
Baseline Low
1
C 110 bpm 120 bpm to 210 bpm, in steps of 10 bpm
Variability Lim. C Guideline Manual, Guideline
Variability High
2
C 25 bpm 5 bpm to 30 bpm, in steps of 1 bpm
Variability Low
2
C 5 bpm 0 bpm to 10 bpm, in steps of 1 bpm
FHR Available C 90 % 50 % to 100%, in steps of 5 %