6 Monitor Settings
83
Configuring Bed Information
Monitor Settings: Main Setup > Bed Information
Bed Information Factory Default Setting
Bed Information Configuration Implications
IP OB Server
Use the IP OB Server setting to define the IP address of the OB server.
This is the IP address of any fetal monitor data acquisition host such as a data acquisition client PC, an
IntelliSpace Perinatal (formerly OB TraceVue) data acquisition server, or a surveillance and archiving
server.
• If the field is left empty (set to 0.0.0.0), the fetal monitor can connect automatically to any available
data acquisition PC in the same subnet.
• If the current host PC is unavailable, the fetal monitor automatically tries to find another host.
• If the IP address of the host PC is entered manually, the fetal monitor can only connect to this
dedicated data acquisition PC, and dynamic hosting is not possible.
Configuring Edit Notes
Monitor Settings: Main Setup > Edit Notes > Edit
Edit Notes Factory Default Settings
Setting Item Mode Factory Default Choice/Range
IP OB Server C 0.0.0.0 n/a
Setting Item Mode Factory Default Choice/Range
Note <1>
1
C Patient Repositioned free configurable, maximum 30 characters
Note <2
1
C Vaginal Examination free configurable, maximum 30 characters
Note <3
1
C MD Notified free configurable, maximum 30 characters
Note <4
1
C Sitting free configurable, maximum 30 characters
Note <5
1
C On Back free configurable, maximum 30 characters
Note <6
1
C Left Lateral free configurable, maximum 30 characters
Note <7
1
C Ambulating free configurable, maximum 30 characters
Note <8
1
C Tocolytic Given free configurable, maximum 30 characters
Note <9>
1
C Membranes Ruptured free configurable, maximum 30 characters
Note <10>
1
C Amniotomy free configurable, maximum 30 characters
Note <11>
1
C Amniotic Fluid Clear free configurable, maximum 30 characters
Note <12>
1
C Amniotic Fluid Not Clear free configurable, maximum 30 characters
Note <13>
1
C Oxytocin free configurable, maximum 30 characters
Note <14>
1
C Urinary Catheter free configurable, maximum 30 characters