5 Measurement Settings
44
NOTE
1
The label iTaxil is currently not supported by the Covidien Genius 2 infrared thermometer.
WARNING
The limit settings can be set differently for each label. Make sure that you configure alarm limit settings
for all possible labels.
Alarms
Use the Alarms setting to switch infrared temperature alarms On or Off. Your monitor must be
configured to
Alarm Mode > All to enable alarms in general.
iTtymp
Use the iTtymp
1
setting to switch the infrared temperature measurement On or Off.
1
The temperature label iTtymp changes automatically depending on the label that is transmitted by the
infrared thermometer.
Label
Use the Label setting to set the alarm limits differently for each infrared temperature label.
The following choices are available:
iTtymp
1
iToral
iTcore
iTaxil
2
iTrect
NOTE
1
Only values labeled with iTtymp are transmitted to a connected IntelliSpace Perinatal system.
2
The label iTaxil is currently not supported by the Covidien Genius 2 infrared thermometer.
Unit
Use the Unit setting to define the unit. Choices are °C and °F.
Color
Use the Color setting to define the infrared temperature measurement color.
Value Lifetime
Use the Value Lifetime setting to define the time after which an aperiodic infrared temperature numeric
value is deleted from the Screen.
Choices are:
30 min, 1 h, 2 h, 4 h, 6 h, 8 h, 12 h, 24 h.