6 Monitor Settings
67
Trace Separation
If the Trace Separation setting is set to On, the recorded FHR traces will be separated from each other
with an offset as configured in the
Separation Order setting.
Choices are
On and Off.
Separation Order
Use the Separation Order setting to define the order of the recorded FHR/DFHR traces if Trace
Separation
is set to On.
•
Separation Order > Standard:
– FHR1 trace is not shifted
– FHR2 trace is shifted 20 bpm up
– FHR3 trace is shifted 20 bpm down
– see the trace below:
•
Separation Order > Classic (was used in Series 50 monitors):
– FHR1 trace is shifted 20 bpm up when other FHR/DFHR measurements are present
– FHR2 trace is not shifted
– FHR3 trace is shifted 20 bpm down
– see the trace below: