6 Monitor Settings
56
Configuring INOP Severity
Monitor Settings: Main Setup > Alarms > INOP Severity
INOP Severity Factory Default Settings
INOP Severity Settings Configuration Implications
ECG Leads Off
Set ECG Leads Off to Yellow or Red if you want the “ECG Leads Off” INOP to be signaled as a yellow or
red INOP. This INOP alerts the clinician when not all required leads for ECG monitoring are attached
to the patient. If, after a discharge, no ECG has been measured yet, the INOP severity will be
Cyan
regardless of the configuration. Once a valid ECG has been received, the severity will behave as
configured.
NOTE
If you reset the fetal monitor configuration with Main Setup > Defaults > Factory Default, this INOP
Severity setting is reset to
Yellow.
Check the settings together with the Fetal Recorder settings.
Battery Empty
Set Battery Empty to Yellow or Red if you want the "Battery Empty" INOP to be signaled as a cyan or red
INOP. This INOP alerts the clinician when a device battery is almost empty and must be replaced.
NOTE
The batteries of the Avalon FM20 and FM30 issue a yellow "Battery Empty" INOP.
The batteries of the Avalon CL transducers and the NBP and SpO2 pods issue a cyan "
Battery
Empty
" INOP.
This setting does not apply for "
Battery Low" INOPs.
If you reset the fetal monitor configuration with
Main Setup > Defaults > Factory Default, this
INOP Severity setting is reset to
Yellow.
Check the settings together with the Fetal Recorder settings.
Cuff Overpress
Set Cuff Overpress to Yellow or Red if you want the “Cuff Overpress” INOP to be signaled as a yellow or
red INOP. This INOP alerts the clinician when the NBP cuff pressure exceeds the overpressure safety
limits.
Setting Item Mode Factory Default Choice/Range
ECG Leads Off C Cyan Cyan, Yellow, Red
Battery Empty C Cyan Cyan, Yellow, Red
Cuff Overpress C Cyan Cyan, Yellow, Red
Cuff NotDeflated C Cyan Cyan, Yellow, Red
No Pulse C Cyan Cyan, Yellow, Red