5 Measurement Settings
28
Configuring HR from MECG
Measurement Settings: Main Setup > Measurements > MECG
HR (MECG) Factory Default Settings
HR (MECG) Configuration Implications
Your monitor must be configured to Alarm Mode > All to enable the MECG/Pulse alarms.
NOTE
If you change one of these settings in the Setup MECG menu, this will also change it in the Setup Pulse
menu and vice versa.
High Limit, Low Limit
HR (MECG) and Pulse (SpO2) share the same alarm limits. These alarm limits apply to the current alarm
source, either
HR or Pulse.
MECG Alarms
Use the MECG Alarms setting to switch the HR alarm On or Off.
NOTE
If MECG is active, Pulse Alarms from SpO2 is grayed out and switched off.
QRS Volume
The QRS Volume setting defines the default volume of the QRS Tone. The QRS Tone is derived from
either the
MECG or Pulse from SpO2, depending on which is currently selected as the Alarm Source.
Setting Item Mode Factory Default Choice/Range
High Limit C, M 120 bpm 31 to 300 bpm
in steps of 1 bpm (31 to 40 bpm)
in steps of 5 bpm (40 to 300 bpm)
Low Limit C, M 50 bpm 30 to 295 bpm
in steps of 1 bpm (30 to 40 bpm)
in steps of 5 bpm (40 to 295)
MECG Alarms C, M On On, Off
QRS Volume C, M This setting is accessible in this menu, but is actually a Monitor Setting,
see “QRS Volume” on page 58.
ΔExtrTachy C 20 bpm 0 to 50 bpm, in steps of 5 bpm
Tachy Clamp C 200 bpm 150 to 240 bpm, in steps of 5 bpm
Δ ExtrBrady C 20 bpm 0 to 50 bpm, in steps of 5 bpm
Brady Clamp C 40 bpm 30 to 100 bpm, in steps of 5 bpm
Color C Red Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red