5 Measurement Settings
26
Configuring Toco
Measurement Settings: Main Setup > Measurements > Toco
Toco Factory Default Settings
Toco Configuration Implications
Set Baseline
This is not a setting. Select Set Baseline to reset the Toco baseline to 20 on the display and the trace.
NOTE
If the Toco value is negative for more that five seconds, the Toco baseline is automatically reset to 0.
Since the lowest value on the scale is
0, the recorder is then again able to print any following
contraction on the paper trace.
Gain
Use the Gain setting to define the Toco sensitivity. If the Toco sensitivity is too high with 100%, and
the Toco trace for example exceeds the paper scale, you can reduce the Toco sensitivity to
50%.
Color
Use the Color setting to define the Toco color.
NOTE
The color settings for Toco and IUP are independent.
Setting Item Mode Factory Default Choice/Range
Set Baseline C, M n/a n/a
Gain C, M 100% 50%, 100%
Color C Green Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red