6 Monitor Settings
52
Red and Yellow Measurement Alarm Latching Behavior
The following table explains the behavior if an alarm has not been acknowledged:
The following table explains the behavior if an alarm has been acknowledged:
Alarm Reminder
Use this setting to define how alarm indications behave if alarm conditions remain active after they
have been acknowledged:
On: After the configured Reminder Time, the alarm tone is repeated for a limited time (6 seconds).
ReAlarm: After the Reminder Time the alarm tone is repeated continuously (this is the same as a
new alarm).
Off: No Alarm Reminder is issued.
Alarm Condition Non-latching alarms Visual and audible
latching
Visual latching,
audible non-latching
Alarm condition still
present
Alarm lamp on
Alarm message visible
Numeric flashing
Alarm tone on
Alarm condition no longer
present
Alarm lamp off
No Alarm message visible
No flashing numeric
Alarm tone off
Alarm lamp on
Alarm message visible
Numeric flashing
Alarm tone on
Alarm lamp on
Alarm message visible
Numeric flashing
Alarm tone off
Alarm Condition Non-latching alarms Visual and audible
latching
Visual latching,
audible non-latching
Alarm condition still
present
Alarm lamp off
Alarm message visible
Numeric flashing
Alarm tone off
Audible Alarm Reminder (if configured)
Alarm condition no longer
present
Alarm lamp off
No Alarm message visible
No flashing numeric
Alarm tone off

Table of Contents