7 Global Settings
86
QRS Type
Use the QRS Type setting to configure the type of the QRS sound which can be high pitched (e.g. for
Japan) or low pitched.
The choices are
QRS Tone and QRS Tick.
For both types, the frequency and rhythm information is derived from either the
MECG/DECG or Pulse,
depending on which is currently active.
If the parameter
Tone Modulation is set to Yes (see “Tone Modulation” on page 58), the QRS Type
automatically switches to
QRS Tone.
Automat. Default
•If Automat. Default is set to Yes, and the monitor is switched off for more than one minute, the
User Defaults are reloaded at the monitor. Any changes that are not stored to the active settings are
lost.
•If
Automat. Default is set to No, and the monitor is switched off for more than one minute, the
active settings from the most recent session are retained.
Automatic Default does not affect the
monitor's behavior when you discharge a patient. After discharge, the
User Defaults are always
restored.
NOTE
If the monitor is switched off and then on again in less than one minute, all active settings are retained,
irrespective of the
Automat. Default setting.
Sensor Disconnct
The default of this setting is No Auto Off.
No sensor INOPs are shown even when alarms are Off or Paused.
In Standby Mode no
unplugged INOPs are generated at all.
•If you set
Sensor Disconnct to Auto Off, and you Confirm the settings change:
When you disconnect the transducer, the parameters are switched off automatically during the
main alarms off state, and the individual alarms off state.
No unplugged INOPs are generated when main alarms are
Off or Paused (or yellow off/yellow
paused
).
Power Loss Sound
FM20, FM30
with #25 only
The Power Loss Sound setting lets you define whether the power loss sound is Enabled or Disabled. If
Enabled, a sound will be generated whenever the main power is lost or the power cord is disconnected
while the monitor is running.
ConfirmAlarmsOff
This ConfirmAlarmsOff setting determines whether pausing alarms, or switching alarms off, has to be
confirmed by the user before it becomes effective. If
ConfirmAlarmsOff is configured to Yes, a pop-up
key line will appear asking to confirm that alarms should be:
1 paused (if Alarms Off setting is set to 1 min, 2 min, or 3 min)
2 switched off (if Alarms Off setting is set to Infinite)

Table of Contents