6 Monitor Settings
78
Recording NST Report Only and Automatically
To configure this use model, configure the following settings:
As soon as the NST trace interpretation is finished or stopped, the recorder prints the
NST Report.
Setting Item Mode
Main Setup > NST Report > Setup > Setup NST Report > Report Recording > Immediately C, M
Main Setup > NST Report > Setup > Setup NST Report > Auto Trace Rec > No C
Main Setup > Fetal Recorder > NST Autostart > Off C
Main Setup > Fetal Recorder > NST Autostop > Off C