6 Monitor Settings
54
IncreaseVolDelay
Use the IncreaseVolDelay setting to define the interval after which the Alarm Volume increases in steps.
Choices are
10 sec, 20 sec, 30 sec.
The following example shows how the settings
Alarm Volume, Red Alarm Volume, Yell.AlarmVolume,
INOP Volume, AutoIncrease Vol, and IncreaseVolDelay work together in case of alarming (using the
factory default settings; see table below):
Coincidence Tone
Immediately after the detection of a coincidence a cyan INOP Coincidence (with tone) is generated.
Use the
Coincidence Tone setting to define the time when a yellow INOP !!Coincidence is generated.
Choices are:
immediately: to generate a yellow INOP !!Coincidence immediately after the coincidence detection
after 30 sec: to generate a yellow INOP !!Coincidence after 30 seconds after the coincidence
detection
after 60 sec: to generate a yellow INOP !!Coincidence after 60 seconds after the coincidence
detection
Setting Item Factory Default
Alarm Volume 5
Red Alarm Volume AlarmVol. +2
Yell.AlarmVolume AlarmVol. +0
INOP Volume AlarmVol. +0
AutoIncrease Vol 2 Steps
IncreaseVolDelay 20 sec