5 Measurement Settings
43
Configuring iTtymp
Measurement Settings: Main Setup > Measurements > iT<label>
1
• iTtymp
• iToral
• iTcore
• iTrect
1
current infrared temperature (iT) label as transmitted by the infrared thermometer.
iTtymp Factory Default Settings
1
This setting is only available if Main Setup > Alarms > Alarm Settings is set to All.
2
The temperature label iTtymp changes automatically depending on the label that is transmitted by the
infrared thermometer.
iTtymp Configuration Implications
High Limit, Low Limit
Use the High Limit and Low Limit settings to configure the following iT temperature labels:
•
iTtymp
• iToral
• iTcore
• iTaxil
1
• iTrect
Choices are:
•from
33.0°C to 35.0°C in steps of 0.5°C
•from
35.1°C to 42.0°C in steps of 0.1 °C
Setting Item Mode Factory Default Choice/Range
High Limit
1
C, M 39.0 36.1 to 42.0 in steps of 0.1
Low Limit
1
C, M 36.0 33.0 to 35.0 in steps of 0.5
35.1 to 38.9 in steps of 0.1
Alarms
1
C On On, Off
iTtymp
2
C Off On, Off
Label C iTtymp itTymp, iToral, iTcore, iTrect
Unit C °C °C, °F
Color C White Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red
Value Lifetime C 1 h 30 min, 1 h, 2 h, 4 h, 6 h, 8 h, 12 h, 24 h
Enter Temp C, M n/a n/a