EasyManua.ls Logo

Philips FM20 - NST Autostart; NST Autostop; Paper Save Mode

Philips FM20
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Monitor Settings
66
NST Autostart
Use the NST Autostart setting to define if the fetal recorder starts when the NST Timer gets started or
not.
Choices are
On and Off.
NST Autostop
Use the NST Autostop setting to define if the fetal recorder automatically stops when the NST Timer has
elapsed or not.
Choices are
On and Off.
Paper Save Mode
Set the Paper Save Mode setting to On to record maternal vital signs using less paper than during normal
trace recording. The
Paper Save Mode recording starts as soon as the recorder is stopped.
Choices are
On and Off.
Every NBP measurement is recorded with a timestamp of the measurement.
•SpO
2
, pulse, and maternal heart rate are recorded every five minutes.
If there is no valid maternal measurement recorded for more than one hour, the recorder stops
and a notifying message is issued.
If a valid measurement is made, the recorder starts again automatically.

Table of Contents

Other manuals for Philips FM20

Related product manuals