6 Monitor Settings
57
Cuff NotDeflated
Set Cuff NotDeflated to Yellow or Red if you want the “Cuff NotDeflated” INOP to be signaled as a
yellow or red INOP. This INOP alerts the clinician when the NBP cuff pressure has exceeded 15
mmHg (2 kPa) for more than 3 minutes.
No Pulse
Set No Pulse to Yellow or Red if you want the INOPs SpO
2
No Pulse to be signaled as a yellow or red
INOP.
Configuring User Interface Settings
Monitor Settings: Main Setup > User Interface
Unique Monitor Settings
: Main Setup > User Interface
User Interface Factory Default Settings
1
QRS Type is a Global Setting, see “QRS Type” on page 86
2
only for FM20/FM30 with battery option
Setting Item Mode Factory Default Choice/Range
QRS Volume C, M 00 to 10
QRS Low C 00 to 10
QRS Type
1
C QRS Tone QRS Tone, QRS Tick
Prompt Volume C 80 to 10
Tone Modulation C, M Yes Yes, No
Tone Mod. Type C Enhanced Standard, Enhanced
Global Speed C, M 25 mm/sec 6.25 mm/sec, 12.5 mm/sec, 25 mm/sec,
50 mm/sec
Touch ToneVolume C 10 to 10
Timer Volume C, M 40 to 10
Global SmartKeys C see “Global SmartKeys
Factory Defaults” on
page 59
see: “Global SmartKeys Choices” on page 60
Brightness C, M Optimum 1 to 10, Optimum
Standby Brightn. C Optimum 1 to 10, Optimum
TransportBrightn
2
C 4 1 to 10, Optimum
Display Units C No Yes, No
Alarm Limits C Yes Yes, No
Measurement Time C Timestamp Timestamp, Time to Next
NBP Sys/Dia only C Yes Yes, No
Wave Line Style C Thin Thin, Medium, Thick, Extra Thick
Aged Numerics C Standard Standard, Grayed Out, Blanked