Recording Trace and NST Report Automatically
This is the factory default behavior of your Avalon fetal monitor.
To configure this use model, configure the following settings:
The real-time trace recording begins upon starting the NST trace interpretation.
As soon as the
NST trace interpretation is finished or stopped, the recorder pauses the real-time
recording to print the
NST Report.
Then the real-time recording continues.
Setting Item Mode
Main Setup > NST Report > Setup > Setup NST Report > Report Recording > Immediately C, M
Main Setup > NST Report > Setup > Setup NST Report > Auto Trace Rec > No C
Main Setup > Fetal Recorder > NST Autostart > On C
Main Setup > Fetal Recorder > NST Autostop > Off C