6 Monitor Settings
69
Configuring NST Timer
Unique Monitor Settings: Main Setup > NST Timer
NST Timer Factory Default Settings
1
On when option #C71 (NST Trace Interpretation) is available
NST Timer Configuration Implications
Run Time
Use the Run Time setting to define the time frame where the NST trace interpretation observes the set
criteria.
Choices are from
10 min to 60 min in steps of 5 min.
Notification
The Notification setting lets you configure an alarm, or a single tone as additional means of notification:
•Select
Alarm to receive an INOP alarm when the timer expires.
•Select
Sound to hear a single tone when the timer expires.
•Select
No Sound for no additional notification.
NOTE
When the Run Time of the NST Timer expires, the notification sound is issued, its color changes from
blue to green, and the timer continues to count until the
NST Report stops.
Timer
Use the Timer setting to switch the NST Timer On or Off.
Choices are
On and Off.
NOTE
•When the NST Timer is switched Off, NST Analysis will also be switched Off.
•When
NST Analysis is switched On, the NST Timer will also be switched On.
•The
Timer setting is actually a Global setting. A setting change does not have to be changed if you
leave the Configuration Mode. It takes effect immediately.
Setting Item Mode Factory Default Choice/Range
Run Time C 20 min 10 min to 60 min, in steps of 5 min
Notification C Sound Alarm, Sound, No Sound
Timer C Off (On
1
)On, Off
Timer Volume C, M 40 to 10