5 Measurement Settings
42
Fast: starts the measurement cycle within a few seconds.
Color
Use the Color setting to define the NBP color.
Configuring Temperature
Global Settings: Main Setup > Measurements > Configure Temp
The manually entered temperature measurement is actually a Global Setting. If you change Global
Settings they do not need to be saved. However, since
Enter Temp appears in the Measurements
window it is listed in this chapter of the Configuration Guide.
To manually enter a temperature value, select
Main Setup > Measurements > Enter Temp.
Temperature Factory Default Settings
1
Entering a manually measured temperature switches the Msmnt automatically On.
Temperature Configuration Implications
Label
The Label setting Temp for manual data entries for temperature cannot be changed.
Unit
Use the Unit setting to define the unit. Choices are °C and °F.
Color
Use the Color setting to define the for the manually entered temperature measurement.
Interval
Use the Interval setting to define the time after which a manually entered temperature value becomes
invalid (no value is then displayed).
Msmnt
Use the Msmnt setting to define whether the selected measurement will be On or Off.
Setting Item Mode Factory Default Choice/Range
Label C Temp n/a
Unit C °C °C, °F
Color C White Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red
Interval C 1 h 30 min, 1 h, 2 h, 3 h, 4 h, 6 h, 8 h, 12 h, 24 h
Msmnt C Off
1
On, Off

Table of Contents