6 Monitor Settings
51
NOTE
In INOP only mode, no patient alarms are enabled or indicated. No alarm limits or alarm off icons are
displayed. No patient alarm settings are available in the setup menus.
Visual Latching, Audible Latching
The alarm latching settings define how the alarm indicators behave when you do not acknowledge
them.
When alarms are set to non-latching, their indicators end when the alarm condition ends.
Switching alarm latching on, means that the visual and/or audible alarm indications are still
displayed or announced after the alarm condition ends. The indications last until you acknowledge
them.
The choices for
Visual Latching and Audible Latching are:
Red&Yellow
Red Only
Off
The choices can be combined to give the following settings:
Visual Latching Audible Latching
Red&Yellow Red&Yellow
Red&Yellow Red Only
Red Only Red Only
Red Only Off
Off Off