6
47
6Monitor Settings
This section lists all the settings grouped in the Monitor Settings block. Read all information on
configuration implications at the end of the relevant tables before you make any configuration
changes.
Understanding Monitor Settings
What sets Monitor Settings apart from Measurement Settings, is that they are not specific to one
measurement. Monitor Settings affect the general behavior of the monitor.
Alarm settings are a good example to help you understand the difference between Monitor and
Measurement Settings: general alarm settings, such as
Alarm Volume or the Alarms Off time are Monitor
Settings. They determine the monitor’s general alarming behavior. Individual alarm limits that can be
set individually for each measurement are Measurement Settings. They only affect the selected
measurement.
Configuring Alarms Settings
Monitor Settings: Main Setup > Alarms > Alarm Settings
Alarms Factory Default Settings
Setting Item Mode Factory Default Choice/Range
Alarm Volume C, M 50 to 10
Alarms Off Prio C Red & Yellow Red & Yellow, Yellow Only, Not Allowed
Alarms Off C 2 min 1 min, 2 min, 3 min, Infinite
Pause Al. 5min C Enabled Enabled, Disabled
Pause Al. 10 min C Enabled Enabled, Disabled
AlarmOffReminder C Off On, Off
AlarmsOffAtStart C No No, Yes
Alarm Mode C INOP only INOP only, All
Visual Latching C Red & Yellow Red & Yellow, Red Only, Off
Audible Latching C Red & Yellow Red & Yellow, Red Only, Off
Alarm Reminder C On On, ReAlarm, Off
Reminder Time C 3 min 1 min, 2 min, 3 min