5 Measurement Settings
27
Configuring IUP
Measurement Settings: Main Setup > Measurements > IUP
IUP Factory Default Settings
IUP Configuration Implications
Set Baseline
This is not a setting. Select Zero IUP to reset the IUP to zero (0) on the display and the trace.
NOTE
If the IUP value is negative for more that five seconds, the IUP baseline is automatically reset to 0.
Since the lowest value on the scale is
0, the recorder is then again able to print any following
contraction on the paper trace.
Unit
Use the Unit setting to define the unit. Choices are mmHg and kPa.
Color
Use the Color setting to define the IUP color.
NOTE
The color settings for Toco and IUP are independent.
Setting Item Mode Factory Default Choice/Range
Zero IUP C, M n/a n/a
Unit C, M mmHg mmHg, kPa
Color C Green Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red