5Measurement Settings
This section lists all the settings grouped in the Measurement Settings block. They define how the
monitor measures and displays patient data. Read all information on configuration implications at the
end of each section before you make any configuration changes.
Configuring FHR/DFHR
Measurement Settings: Main Setup > Measurements > FHR <x>
Measurement Settings: Main Setup
> Measurements > DFHR <x>
FHR<n>, DFHR<n> Factory Default Settings
1
in the Setup FHR <x> menu only
2
in the Setup DFHR <x> menu only
Setting Item Mode Factory Default Choice/Range
Select Audio C, M n/a n/a
Audio Select C On Connect On Connect, Manual Only
FHR Volume C, M 60 to 10
High Limit C, M 150 70 .to 210 bmp, in steps of 10 bmp
Low Limit C, M 110 60 to 200 bmp, in steps of 10 bmp
Alarms C, M On On, Off
Trace Separation C, M Off On, Off
Fetal Movement
1
C, M On On, Off
ArtifactSuppress
2
C, M On On, Off
DECG Wave
2
C, M Off On, Off
High Delay C, M 60 sec 10 to 300 sec, in steps of 10 second
Low Delay C, M 60 sec 10 to 300 sec, in steps of 10 second
SignalLoss Delay C, M 60 sec 10 to 300 sec, in steps of 10 second
Color C Orange Red, Green, Yellow, Blue, Magenta, Cyan,
White, Pink, Orange, Light Green, Light Red