7Global Settings
This section lists all Global Settings. Global Settings are set once per monitor, and are independent of
the
Profiles and settings blocks (see “Understand the Configurable Software Elements” on page 9). The
difference is that any changes you may configure are automatically stored, there is no need to save
them. Global Settings can be cloned from a monitor's configuration (cfg.) file.
Read all information on configuration implications at the end of the sections before you make any
configuration changes.
Configuring General Global Settings
Global Settings: Main Setup > Global Settings
Global Settings Factory Default Settings
1
geography-specific factory default setting
2
FM20/FM30 with battery option (#E25) only
Global Settings Configuration Implications
Line Frequency
Use the Line Frequency setting to configure the correct line frequency for the AC power, either 50 Hz
or
60 Hz. If the Line Frequency is set incorrectly, it may affect the ECG signal quality.
Setting Item Mode Factory Default Choice/Range
Line Frequency
1
C 50 Hz <or> 60 Hz
1
50 Hz, 60 Hz
QRS Type
1
C QRS Tone <or> QRS
Tick
1
QRS Tone, QRS Tick
Automat. Default C Yes Yes, No
Sensor Disconnct C No Auto Off Auto Off, No Auto Off
Power Loss Sound
2
C Enabled Enabled, Disabled
ConfirmAlarmsOff C No Yes, No
LAN Data Export C All All, Anonymous, Off