6 Monitor Settings
48
Alarms Configuration Implications
Alarm Volume
Use the Alarm Volume setting to define the base volume of the red and yellow audible alarm indicators
and the INOP tones.
Alarms Off Prio
The setting Alarms Off Prio defines the level of alarms that are suppressed in case main alarms are
switched off or paused via:
Alarms Off (if the Alarms Off setting is set to Infinite):
–via Smart Key
–via
Alarm Settings menu operation
Pause Alarms (if the Alarms Off setting is set to 1 min, 2 min, 3 min):
–via Smart Key
–via
Alarm Settings menu operation
Possible choices are:
Red & Yellow
Yellow Only
Not Allowed
Alarm Sounds C Traditional Traditional, ISO
RedAlarmInterval C 10 sec 5 sec, 10 sec, 15 sec
Yel.Al. Interval C 20 sec 10 sec, 20 sec, 30 sec
Alarm Low C 40 to 10
Red Alarm Volume C AlarmVol. +2 AlarmVol. +0, AlarmVol. +1, AlarmVol. +2
Yell.AlarmVolume C AlarmVol. +0 AlarmVol. +0, AlarmVol. +1, AlarmVol. +2
INOP Volume C AlarmVol. +0 AlarmVol. +0, AlarmVol. +1, AlarmVol. +2
AutoIncrease Vol C 2 Steps Off, 1 Step, 2 Steps
IncreaseVolDelay C 20 sec 10 sec, 20 sec, 30 sec
Coincidence Tone C immediately immediately, after 30 sec, after 60 sec
Keep Blinking C No Yes, No
Relay1 Sensitiv. C R&Y&C Red, Red&Yell, Red&Cyan, R&Y&C
Relay2 Sensitiv. C Red & Yell Red, Red&Yell, Red&Cyan, R&Y&C
Relay3 Sensitiv. C Red Red, Red&Yell, Red&Cyan, R&Y&C
CyanRelayLatency C 0 sec 0 sec, 2 sec, 5 sec
Yel.RelayLatency C 5 sec 0 sec, 2 sec, 5 sec, 10 sec
Alarm Text C Standard Standard, Enhanced
Setting Item Mode Factory Default Choice/Range