6 Monitor Settings
49
The following table applies:
1
This setting is only available if Main Setup > Alarms > Alarm Settings is set to All.
The Alarms Off/Pause Alarms SmartKey, and the alarm flag on the Screen change their text
automatically and accordingly as following:
Setting Alarm Type Visual Alarm Audible Alarm
Red & Yellow Red Alarm no no
Yellow Alarm
no no
Short Yellow Alarm
no no
Red INOP
yes no
Yellow INOP
yes no
Hard INOP
yes no
Soft INOP
yes no
Yellow Only
1
Red Alarm yes yes
Yellow Alarm
no no
Short Yellow Alarm
no no
Red INOP
yes yes
Yellow INOP
yes no
Hard INOP
yes no
Soft INOP
yes no
Not Allowed No possibility to turn main alarms Off.