3 Configuring Active and Default Settings
14
In Configuration Mode only, these can be loaded into the Active Settings with:
Main Setup > Defaults > Factory Default
You can use the Factory Defaults as the basis for producing your User Defaults.
CAUTION
This resets all settings to factory defined values, but be aware that some values will differ from those
with which the fetal monitor was originally shipped from the factory After loading the Factory
Defaults, check the settings, and if necessary, change them to the settings you normally use.
Some settings will be set to <Unknown>. Make sure to change the following setting to the setting
choices/ranges you want to use:
Recorder Speed
Scale Type
Frequency Band
Channel
The Battery Empty INOP Severity setting will be set to Yellow. Make sure to change it back to Cyan if
this was the user's choice.