10 Instrument setup
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Instrument setup 191
Configuration Database > Demographics
Configuration Lab > Sample Types
If sample type is enabled in Configuration Processing
> Result Output, this Sample Type Short Name is
used on the host interface.
q If sample-type specific racks are not used, this
Sample Type Short Name has to be included in the
barcode.
Parameter Value
Patient ID Selected All sample orders have to be
assigned to an existing patient,
i.e. prior to downloading
orders, the relevant patient
information has to be
downloaded. The instrument
generally works with patient
data.
Not selected The instrument generally
works without patient data.
Orders have to be entered
without patient data.
Free Text
Fields
Selected The instrument allows defining
laboratory specific fields. The
fields and their assigned
names are shown in the
patient or order menu. The
relevant information has to be
transmitted via the host
interface.
Not selected The free text fields will not be
shown in the patient or order
menu. On the host interface,
the relevant fields within the
Patient and Order Entry
blocks may be left empty.
y Database parameters