Remote Control Commands
R&S
®
RTB2000
406User Manual 1333.1611.02 ─ 04
Parameters:
<DataLength> Range:  0  to  8
Increment:  1
*RST:  1
Default unit: Byte
TRIGger:A:CAN:DATA <Data>
Defines the data pattern. The number of bytes in the data pattern is defined with
TRIGger:A:CAN:DLC.
The command is relevant if TRIGger:A:CAN:TYPE is set to IDDT.
Parameters:
<Data> String containing binary pattern with max. 64 bit. Characters 0,
1, and X are allowed. Make sure to enter complete bytes.
TRIGger:A:CAN:ACKerror <AcknowledgeError>
Triggers on acknowledgement errors. An acknowledgement error occurs when the
transmitter does not receive an acknowledgment - a dominant bit during the Ack Slot.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Parameters:
<AcknowledgeError> ON | OFF
*RST:  OFF
TRIGger:A:CAN:BITSterror <BitStuffingError>
Triggers on bit stuffing errors.
See also: "Stuff Bit" on page 199.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Parameters:
<BitStuffingError> ON | OFF
*RST:  ON
TRIGger:A:CAN:CRCerror <CRCerror>
Triggers on errors in the Cyclic Redundancy Check.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Parameters:
<CRCerror> ON | OFF
*RST:  OFF
Serial Bus Analysis