EasyManua.ls Logo

Rohde & Schwarz R&S RTB2004 - Page 443

Rohde & Schwarz R&S RTB2004
509 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control Commands
R&S
®
RTB2000
435User Manual 1333.1611.02 ─ 04
To set the range of samples to be returned, use DIGital<m>:DATA:POINts.
Suffix:
<m>
.
0..15
Selects the logic channel
<p> 1..2
Selects the logic pod
Parameters:
<WaveformData> List of values according to the format settings.
Example:
FORM ASC,0
DIG1:DATA?
1,1,1,1,1,1,0,0,0,0,0,0,...
Usage: Query only
LOGic<p>:DATA:HEADer?
DIGital<m>:DATA:HEADer?
Returns information on the waveform of the specified logic channel/pod.
Table 15-6: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 200000
4 Number of values per sample interval, usually 1. 1
Suffix:
<m>
.
0..15
Selects the logic channel
<p> 1..2
Selects the logic pod
Parameters:
<Header> Comma-separated value list
Example: -9.477E-008,9.477E-008,200000,1
Usage: Query only
LOGic<p>:DATA:POINts <PointSelection>
DIGital<m>:DATA:POINts <PointSelection>
As a setting, the command selects a range of samples that will be returned with
DIGital<m>:DATA?. As a query, it returns the number of returned samples for the
selected range.