Remote Control
R&S
®
RTM20xx
304User Manual 1317.4726.02 ─ 01
TRIGger:A:HYSTeresis <Hysteresis>
Sets a hysteresis range around the trigger level of the A trigger event. If the signal jitters
inside this range and crosses the trigger level thereby, no trigger event occurs.
Hysteresis is available for edge and width trigger.
Parameters:
<Hysteresis> AUTO | SMALl | MEDium | LARGE
*RST:  AUTO
TRIGger:A:HOLDoff:MODE <HoldOffMode>
Enables or disables the holdoff time.
Parameters:
<HoldOffMode> TIME | OFF
*RST:  Off
TRIGger:A:HOLDoff:TIME <HoldOffTime>
Defines the holdoff time. The next trigger occurs only after the holdoff time has passed.
Parameters:
<HoldOffTime> Default unit: s
15.2.5.2 Edge Trigger
TRIGger:A:EDGE:SLOPe...............................................................................................304
TRIGger:A:EDGE:COUPling...........................................................................................305
TRIGger:A:EDGE:FILTer:LPASs.....................................................................................305
TRIGger:A:EDGE:FILTer:NREJect...................................................................................305
TRIGger:A:EDGE:SLOPe <Slope>
Sets the slope for the edge trigger (A trigger).
Parameters:
<Slope> POSitive | NEGative | EITHer
POSitive
Rising edge, a positive voltage change
NEGative
Falling edge, a negative voltage change
EITHer
Rising as well as the falling edge
*RST:  POSitive
 
Command Reference