Remote Control
R&S
®
RTM20xx
312User Manual 1317.4726.02 ─ 01
TRIGger:A:RISetime:DELTa <Variation>
Sets a time range around the time value defined using TRIGger:A:RISetime:TIME if
TRIGger:A:RISetime:RANGe is set to WITHin | OUTSide.
Parameters:
<Variation> Default unit: s
15.2.5.7 B-Trigger
TRIGger:B:ENABle.........................................................................................................312
TRIGger:B:SOURce.......................................................................................................312
TRIGger:B:EDGE:SLOPe...............................................................................................312
TRIGger:B:LEVel...........................................................................................................312
TRIGger:B:FINDlevel......................................................................................................313
TRIGger:B:MODE..........................................................................................................313
TRIGger:B:DELay..........................................................................................................313
TRIGger:B:EVENt:COUNt...............................................................................................313
TRIGger:B:HYSTeresis...................................................................................................314
TRIGger:B:ENABle <State>
Activates or deactivates the second trigger. The instrument triggers if both trigger event
conditions (A and B) are fulfilled.
Parameters:
<State> ON | OFF
*RST: OFF
TRIGger:B:SOURce <Source>
Selects one of the input channels as B-trigger source. Available channels depend on the
instrument type.
Parameters:
<Source> CH1 | CH2 | CH3 | CH4
*RST: CH1
TRIGger:B:EDGE:SLOPe <Slope>
Sets the edge for the B-trigger.
Parameters:
<Slope> POSitive | NEGative | EITHer
*RST: POSitive
TRIGger:B:LEVel <Level>
Sets the trigger level for the B-trigger event.