EasyManua.ls Logo

Rohde & Schwarz RTM2032 - Page 425

Rohde & Schwarz RTM2032
499 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Remote Control
R&S
®
RTM20xx
410User Manual 1317.4726.02 ─ 01
TRIGger:A:CAN:DATA <Data>
Defines the data pattern. The number of bytes in the data pattern is defined with
TRIGger:A:CAN:DLC.
The command is relevant if TRIGger:A:CAN:TYPE is set to IDDT.
Parameters:
<Data> String containing binary pattern with max. 64 bit. Characters 0, 1,
and X are allowed. Make sure to enter complete bytes.
TRIGger:
A:CAN:ACKerror <AcknowledgeError>
Triggers on acknowledgement errors. An acknowledgement error occurs when the trans-
mitter does not receive an acknowledgment - a dominant bit during the Ack Slot.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Parameters:
<AcknowledgeError> ON | OFF
*RST: OFF
TRIGger:A:CAN:BITSterror <BitStuffingError>
Triggers on bit stuffing errors.
See also: "Stuff bit" on page 188.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Parameters:
<BitStuffingError> ON | OFF
*RST: ON
TRIGger:A:CAN:CRCerror <CRCerror>
Triggers on errors in the Cyclic Redundancy Check.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Parameters:
<CRCerror> ON | OFF
*RST: OFF
TRIGger:A:CAN:FORMerror <FormError>
Triggers on form errors. A form error occurs when a fixed-form bit field contains one or
more illegal bits.
The command is relevant if TRIGger:A:CAN:TYPE is set to ERRCondition.
Command Reference