326Operating Manual 1411.4060.32 ─ 09
[:SOURce<hw>]:PULM:DOUBle:DELay <Delay>
Sets the delay from the start of the first pulse to the start of the second pulse.
Parameters:
<Delay> float
Range: 1 us to 1 s
Increment: 100 ns
*RST: 700 us
Example:
PULM:DOUB:DEL 22 us
22 us elapse between the beginning of the first pulse and the
beginning of the second pulse in double-pulse mode.
Manual operation: See " Double Pulse Delay - Pulse Generator " on page 188
[:SOURce<hw>]:PULM:DOUBle:STATe <State>
Enables/disables double pulse generation. The two pulses are generated in one pulse
period.
Parameters:
<State> 0 | 1 | OFF | ON
*RST: 0
Example:
PULM:DOUB:STAT ON
double-pulse mode is enabled.
[:SOURce<hw>]:PULM:DOUBle:WIDTh <Width>
Sets the width of the second pulse in case of double pulse generation.
Parameters:
<Width> float
Range: 1 us to 1 s
Increment: 100 ns
*RST: 200 us
Example:
PULM:DOUB:WIDT 33 us
sets a width of 33 us for the second pulse.
Manual operation: See " Double Pulse Width - Pulse Generator " on page 188
[:SOURce<hw>]:PULM:MODE <Mode>
Sets the mode of the pulse generator.
SOURce Subsystem