Chapter 9. Probes 9 17
9.1.5.2. Cleaning the probe
The probe must be cleaned and disinfected after each use. Cleaning is an important procedure
that must be carried out before disinfecting the probe.
CAUTION
Do not use brushes when cleaning probes. Even the use of soft brushes can
damage the probe.
During cleaning and disinfection, keep the parts of the probe that must remain
dry higher than the other parts, until all parts are dry.
1. Disconnect the probe from the system.
2. Remove the biopsy adapter or needle guide. Collect the reusable parts, reusable biopsy
adapter or needle guide for disinfection or sterilization and discard the disposable parts in
the clinical waste bin (For cleaning and disinfection/sterilization of the reusable parts please
refer to the appropriate chapter).
3. Remove and properly dispose the medical sheath in the clinical waste bin (Medical sheaths
are disposables).
4. Use a soft cloth, lightly dampened with mild soap or compatible cleaning solution or
compatible cleaning sheet, to remove any particulate matter and bodily fluids that remained
on the probe, edges and ridges on the probe and cable.
5. Rinse with water to remove any loose particles, soap- and cleaning residue (up to the
immersion point).
6. Wipe with a clean, dry soft cloth.
7. If still necessary, wipe with a clean, water-dampened cloth to remove soap residue, and wipe
with a clean, dry soft cloth again.
8. Dispose the used soft cloth in the clinical waste bin.