9 – 18 HS40 Service Manual
9.1.5.3. Disinfecting the probe
The probe must be cleaned and disinfected after each use. Cleaning is an important
procedure that must be carried out before disinfecting the probe. For cleaning the probe
please refer to the appropriate chapter. Disinfect the probe to reduce pathogens to the level
of 10
-6
.
■ Disinfecting the probe with a fluid disinfectant
Using an inappropriate disinfectant may damage the probe.
WARNING
Always use protective eyewear and gloves when disinfecting probes.
Inspect the housing, strain relief, lens and seal for damage, and check for any
functional problem after disinfecting the probe.
Never use disinfectants after the expiration date.
The type of tissue it will contact during use dictates the level of disinfection
required for a device. Ensure that the solution strength and duration of
contact are appropriate for disinfection.
CAUTION
Using a non-recommended disinfectant or not following the recommended
disinfection method can damage and/or discolor the probe, and will void the
probe warranty.
Do not immerse probes for longer than one hour, unless they are sterilizable.
Only sterilize probes by using liquid solutions. Avoid using autoclave, gas
(EtO), or other non-Samsung Medison-approved methods.