EasyManua.ls Logo

Samsung HS40 - Page 267

Samsung HS40
308 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9. Probes 9 19
1. Follow the instructions on the disinfectant label for storage, use and disposal of the
disinfectant.
2. Mix the disinfectant compatible with your probe according to the instructions for solution
strength provided on the disinfectants label.
3. Immerse the probe into the disinfectant as shown in the illustration below.
[Figure 9.2 Disinfection]
4. Follow the instructions on the disinfectant to complete the immersion procedure of the
probe, never leave the transducer longer than the maximum immersion duration in the
fluid. Rinse the probe with sterile water afterwards.
5. Air dry the probe and store it properly (e.g. in a sterile bag), so that no new contamination
may appear.

Table of Contents