9 20 HS40 Service Manual
Disinfecting the probe with a disinfection wipe
Using an inappropriate disinfection wipe may damage the probe.
WARNING
Always use protective eyewear and gloves when disinfecting probes.
Inspect the housing, strain relief, lens and seal for damage, and check for any
functional problem after disinfecting the probe.
Never use disinfection wipes after the expiration date.
Ensure that the contact duration is appropriate for disinfection.
CAUTION
Using a non
-recommended disinfection wipe or not following the recommended
disinfection method can damage and/or discolor the probe. This could also void the probe
warranty.
1. Refer to the user instructions of the disinfection wipe for details of proper use, storage
and disposal of the holder.
2. Dependent on the type of wipe the active disinfectant needs to be applied to a dry wipe or
has already been applied to the wipe.
3. Use the disinfection wipe to spread the active disinfectant over the full surface of the
probe and ensure all areas are covered.
4. Discard the disinfection wipe to clinical waste. Do not re-use the disinfection wipe.
5. Air dry the probe as described in the user instructions of the disinfection wipe for the
described time and store the probe properly (e.g. in a sterile bag), so that no new
contamination may appear.