Chapter 9. Probes 9 – 21
■ Disinfecting the probe with a disinfection system
Using a non-approved probe/disinfection system combination may damage the probe.
WARNING
Always use protective eyewear and gloves when disinfecting probes.
Inspect the housing, strain relief, lens and seal for damage, and check for any
functional problem after disinfecting the probe.
Ensure that the cycle duration is appropriate for the desired disinfection level.
-approved probe/disinfection system combination or not following the
recommended disinfection method can damage the probe. This could also void the probe
warranty.
1. Refer to the user instructions of the disinfection device for details of proper use, storage
and disposal of waste.
2. Place the probe head, a part of the cable and possible marker in the disinfection device,
following the user instructions of the disinfection device.
3. Start the desired disinfection process and wait until the end of the cycle.
4. Remove the probe and possible marker at the end of the cycle from the disinfection
device and store the probe properly (e.g. in a sterile bag), so that no new contamination
may appear.
5. Register the disinfection data for traceability according to local regulations.