Operating Instructions Cubis
®
MCA 87
Operation
8.14 Running Applications (Examples)
8.14.1 Executing the “Unit Change” Function
The “Unit change” function enables the user to switch between the
different units and resolutions defined in the weighing profile of the active
task. The units and resolutions can be set at the beginning of the
weighingprocess.
Procedure
t Start the desired task.
t Press the [Unit change] button.
y All units defined in the weighing profile for the active task are displayed
in a list.
y All resolutions for the weight value defined in the weighing profile for
theactive task are displayed in a list.
t Press the desired unit.
t To set the resolution for the selected unit: Press the desired resolution.
t To confirm the selection and return to the weighing display: Press the
[OK] button.
y The current weight value is displayed in the selected unit and resolution.
8.14.2 Running the “Statistics” Application
(from QAPP Package “Essentials”)
The “Statistics” application saves up to 100 weight values and evaluates
these statistically.
The following values are saved and exported by the statistics application:
− Number of components
− Mean value
− Standard deviation
− Variation coefficient
− Sum of all values
− Lowest value (minimum)
− Highest value (maximum)
− Spread: Difference between maximum and minimum
The “Statistics” application can be combined with the following functions:
− Unit change, only available in the weighing display, before saving
theinitial weight value
− ID marking
− Automatic taring