EasyManua.ls Logo

Sartorius Cubis MCA2.7S - Integrated Clock; Backup Battery; Metrological Data; Models MCA2.7 S | MCA10.6 S | MCA6.6 S | MCA3.6 P

Sartorius Cubis MCA2.7S
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
100 Operating Instructions Cubis
®
MCA
Technical Data
15.5 Integrated Clock
Unit Value
Maximum deviation per month (RTC) s 30
15.6 Backup Battery
Unit Value
Lithium battery, type CR2032
Service life at room temperature, minimum Years 10
15.7 Metrological Data
15.7.1 Models MCA2.7S | MCA10.6S | MCA6.6S | MCA3.6P
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
Scale interval (d) µg 0.1 1 1 1 | 2 | 5
Maximum capacity (Max) g 2.1 10.1 6.1 1.1 | 2.1 | 3.1
Repeatability at 5% load
Standard deviation of the load values,
tolerance
µg 0.20 1 1 3
Standard deviation of the load values,
typical value
µg 0.15 0.4 0.4 0.4
Repeatability at approx. maximum
capacity
Standard deviation of the load values,
tolerance
µg 0.25 1 1 5
Standard deviation of the load values,
typical value
µg 0.18 0.6 0.6 0.6
Linearity deviation
Tolerance µg 0.9 4 4 4
Typical value µg 0.7 3 3 3
Deviation when load is off-center,
positions according to OIML R76
Test weight g 1 5 2 1
Tolerance µg 0.7 4 4 5
Typical value µg 0.5 3 3 3

Table of Contents