Operating Instructions Cubis
®
MCA 57
System Settings
7.7.4 Parameters in the “Weighing and Print Profile / Weighing” Menu
Parameter Setting Values Explanation
Ambient
conditions
Very stable Sets the ambient conditions to “very stable”: Activates a fast
change in the weight values in the event of a load change with
ahigh output rate.
Recommended for the following work environment:
− Very stable table near the wall
− Closed and calm room
Stable Sets the ambient conditions to “stable”.
Recommended for the following work environment:
− Stable table
− Slight movement in the room
− Slight draft
Unstable* Sets the ambient conditions to “unstable”: Activates the delayed
change in weight values with a reduced output rate.
Recommended for the following work environment:
− Simple office desk
− Room with moving machinery or personnel
− Slight air movement
Very unstable Sets the ambient conditions to “very unstable”: Activates a
significantly delayed change in the weight values and long wait for
stability with a further reduction in the output rate.
Recommended for the following work environment:
− Noticeable and slow floor vibrations
− Noticeable building vibrations
− Weighed goods moved
− Very strong air movements
Application filter Weighing (final
readout)*
Activates a filter that enables a fast change in the display for very
fast load changes.
Display changes with minimal load changes (in the digit range)
occur more slowly.
Dosing (filling mode) Activates a filter that enables a very fast change in the display with
minimal load changes (e.g. when dosing or filling vessels).
Sensor mode
(lowfilter)
Activates a weak but fast filter that always behaves in the same
way for load changes (e.g. when dosing in automated systems).
Dynamic mode
(without filter)
Deactivates the active application filter.
Stability Very high accuracy Sets the stability to “very high accuracy”.
High accuracy Sets the stability to “high accuracy”.
Medium accuracy* Sets the stability to “medium accuracy”.
Fast Sets the stability to “fast”.
Very fast Sets the stability to “very fast”.
Very slow Sets the stability to “very slow”.
*Factory setting for the “Standard Weighing” factory-set weighing profile