Technical Data
Unit Value
Pollution level according to IEC 61010-1 2
Overvoltage category according to IEC 606641-1 II
Installation site according to IEC 62368-1, maximum altitude above sea level m 5000
Temperature
In operation °C 0 – +40
During storage and transport °C -20 – +80
Connector for mains connection according to IEC 60320-1 / C14: 3-pin
Power supply cable
Power supply cable according to IEC 60320-1 / C14:
Country-specific, 3-pin, two-sided plug
Other data: See label on AC adapter
15.2.3 Safety of Electrical Equipment
According to EN 61010-1 / IEC 61010-1 Safety requirements for electrical
equipment for measurement, control, and laboratory use –
Part 1: GeneralRequirements
15.2.4 Electromagnetic Compatibility
Interference resistance
Suitable for use in industrial areas
Transient emissions
Class B
Suitable for use in residential areas and areas that are connected to a low
voltage network that also supplies residential buildings.
15.3 Ambient Conditions
15.3.1 Installation Site
Unit Value
Installation site
Standard laboratory rooms
Installation site according to IEC 60259-1, maximum altitude above sea level m 3000
For indoor use only