76 Operating Instructions Cubis
®
MCA
System Settings
7.7.29 Parameters in the “Device Settings / Sound (Loudspeaker)” Menu
Parameter Setting Values Explanation
Touch sounds Activated Deactivates the acoustic signal for touch and keypad operation.
Deactivated* Activates the acoustic signal for touch and keypad operation.
Message sounds Activated Deactivates the acoustic signal for messages.
Deactivated* Activates the acoustic signal for messages.
Sound for
end-of-action
Activated Deactivates the acoustic signal for the completion of an action.
Deactivated* Activates the acoustic signal for the completion of an action.
* Factory setting
7.7.30 Parameters in the “Device Maintenance” Menu
Parameter Setting Values Explanation
Update firmware List of connectors and
then list of firmware
Starts a firmware update.
Not for conformity-assessed devices.
Install
QAPPCenter
List of connectors
andthen list of
QAPPCenters
Updates the QAPP Center using the selected connector.
Install customized
QAPP
List of connectors and
then list of customized
QAPPs
Updates a customized QAPP using the selected connector.
Device
databackup
List of connectors Saves the device data on the selected connector.
Restore
devicedata
List of connectors Loads saved device data from the selected connector.
Export device data Role management Saves device role management for an administrator, user, or guest
on the selected connector.
User management Saves device user management for an administrator, user, or guest
on the selected connector.
Export overview Displays the list of connectors, e.g., a server or connected USB
stick.
Import device data List of connectors
forimport
Loads the device data with role and user management, e.g., from
aserver or connected USB stick.
Reset to factory
settings
Restore factory
settings
Resets the device to factory settings and deletes all
configurations.
The audit trail and data memory will be retained.
Create
diagnosticfile
Saves information about device configuration and operation on
aconnected USB stick.