Operating Instructions Cubis
®
 MCA  71
    System Settings 
Parameter Setting Values Explanation
Safety level Off Deactivates the message for leveling or executing isoCAL at the start 
of the task. The execution of the leveling or isoCAL function can 
betriggered manually via the respective button in the display.
Warning* If the isoCAL function or the leveling must be carried out: The device 
displays a message. Some device functions are limited until isoCAL 
or the leveling has been executed. The message remains intact in the 
event of an abort. The execution of the leveling or isoCAL function 
can be triggered manually via the respective button in the display. 
The displayed weight value is marked as invalid; printed data are 
marked with [!].
Strict If the isoCAL function or the leveling must be carried out: The device 
displays a message. Some device functions are limited until isoCAL 
or the leveling has been executed. Upon aborting the function, the 
task is not started; the function for the message must be executed. 
Ifa message appears while executing the task, the weighed value 
ismarked as invalid. The takeover of this weighed value is blocked 
automatically. The weighed value can be taken over again only when 
the function for message has been executed. The data output is 
deactivated.
* Factory setting
7.7.20  Parameters in the “Device Settings / Preload” Menu
Parameter Setting Values Explanation
Preload Set preload  − Sets the load present on the device as a preload and therefore 
changes the zero point of the device.
 − Reduces the maximum capacity by the set preload.
Delete preload  − Deletes the set preload.
 − Changes the zero point and maximum capacity back to the 
values they were before the preload was set.
7.7.21  Parameters in the “Device Settings / Calibration / Adjustment Report” Menu
Parameter Setting Values Explanation
Automatic 
printout for 
isoCAL
On Activates the automatic printout for the “isoCAL” function.
Off* Deactivates the automatic printout for the “isoCAL” function.
Automatic 
printout for 
cal./adj. task
On Activates the automatic printout for the functions of the 
“calibration/adjustment” tasks.
Off* Deactivates the automatic printout for the functions of the 
“calibration/adjustment” tasks.
Print profile List of print profiles Selection of a print profile via which the printout must be 
carriedout.
* Factory setting