Operating Instructions Cubis
®
MCA 71
System Settings
Parameter Setting Values Explanation
Safety level Off Deactivates the message for leveling or executing isoCAL at the start
of the task. The execution of the leveling or isoCAL function can
betriggered manually via the respective button in the display.
Warning* If the isoCAL function or the leveling must be carried out: The device
displays a message. Some device functions are limited until isoCAL
or the leveling has been executed. The message remains intact in the
event of an abort. The execution of the leveling or isoCAL function
can be triggered manually via the respective button in the display.
The displayed weight value is marked as invalid; printed data are
marked with [!].
Strict If the isoCAL function or the leveling must be carried out: The device
displays a message. Some device functions are limited until isoCAL
or the leveling has been executed. Upon aborting the function, the
task is not started; the function for the message must be executed.
Ifa message appears while executing the task, the weighed value
ismarked as invalid. The takeover of this weighed value is blocked
automatically. The weighed value can be taken over again only when
the function for message has been executed. The data output is
deactivated.
* Factory setting
7.7.20 Parameters in the “Device Settings / Preload” Menu
Parameter Setting Values Explanation
Preload Set preload − Sets the load present on the device as a preload and therefore
changes the zero point of the device.
− Reduces the maximum capacity by the set preload.
Delete preload − Deletes the set preload.
− Changes the zero point and maximum capacity back to the
values they were before the preload was set.
7.7.21 Parameters in the “Device Settings / Calibration / Adjustment Report” Menu
Parameter Setting Values Explanation
Automatic
printout for
isoCAL
On Activates the automatic printout for the “isoCAL” function.
Off* Deactivates the automatic printout for the “isoCAL” function.
Automatic
printout for
cal./adj. task
On Activates the automatic printout for the functions of the
“calibration/adjustment” tasks.
Off* Deactivates the automatic printout for the functions of the
“calibration/adjustment” tasks.
Print profile List of print profiles Selection of a print profile via which the printout must be
carriedout.
* Factory setting