Page #92 background image
92 Operating Instructions Cubis
®
MCA
Malfunctions
10 Malfunctions
10.1 Status Messages
Status
Message
Fault Cause Remedy Chapter,
page
The LED strip
on the
weighing
module is
illuminated
in orange
The device displays
an error message.
The device has detected
a fault.
Rectify the error shown in
theoperating display.
10.2 Warning Messages
Warning
message
Fault Cause Remedy Chapter,
page
Disp.Err. The value to be
output cannot be
shown in the
operating display.
The data to be displayed
is not compatible with the
set display format.
Adjust the display settings
inthe menu, e.g. resolution,
unit, decimal places.
High The device is
overloaded.
The device’s maximum
weighing capacity has
been exceeded.
Reduce the applied weight to
below the device’s maximum
weighing capacity.
15.7,
100
Low The modulation
ofthe weighing
module’s weighing
converter is too low.
No weighing pan has
been placed on the
balance.
A previously forgotten
weight was removed after
starting the device.
Insert the weighing pan into
the device and switch the
device off and on again.
Com.Err. The device is not
receiving any weight
values.
No communication exists
between the display and
control unit and the
weighing module.
Wait until the display and
control unit restores the
communication with the
weighing module.
If the problem occurs again:
Contact Sartorius Service.
17, 106