Operating Instructions Cubis
®
MCA 101
Technical Data
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
Sensitivity drift between +10°C and +30°C ppm/K 1 1 1 1
Tare maximum capacity:
Less than 100% of maximum capacity
Accuracy class according
toDirective2014/31/EU
I I I I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according
toDirective 2014/31/EU
mg 0.01 0.1 0.1 0.1
Minimum initial weighing according
toUSP (United States Pharmacopeia),
Chap. 41
Optimum minimum initial weighing µg 82 820 820 820
Typical minimum initial weighing µg 300 820 820 820
Typical stabilization time s 7 5 5 5
Typical measurement time s 10 8 8 8
15.8 Recommended Calibration Weight
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
External test weight g 2 10 5 3
Recommended accuracy class E2 E2 E2 E2
15.9 isoCAL Function
Unit Value
isoCAL is triggered
bythefollowingcriteria:
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling