Operating Instructions Cubis
®
MCA 101
Technical Data
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
Sensitivity drift between +10°C and +30°C ppm/K 1 1 1 1
Tare maximum capacity:
Less than 100% of maximum capacity
Accuracy class according
toDirective2014/31/EU
I I I I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according
toDirective 2014/31/EU
mg 0.01 0.1 0.1 0.1
Minimum initial weighing according
toUSP (United States Pharmacopeia),
Chap. 41
Optimum minimum initial weighing µg 82 820 820 820
Typical minimum initial weighing µg 300 820 820 820
Typical stabilization time s 7 5 5 5
Typical measurement time s 10 8 8 8
15.8 Recommended Calibration Weight
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
External test weight g 2 10 5 3
Recommended accuracy class E2 E2 E2 E2
15.9 isoCAL Function
Unit Value
isoCAL is triggered
bythefollowingcriteria:
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling