EasyManuals Logo

Sartorius Cubis MCA2.7S User Manual

Sartorius Cubis MCA2.7S
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #101 background imageLoading...
Page #101 background image
Operating Instructions Cubis
®
MCA 101
Technical Data
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
Sensitivity drift between +10°C and +30°C ppm/K 1 1 1 1
Tare maximum capacity:
Less than 100% of maximum capacity
Accuracy class according
toDirective2014/31/EU
I I I I
Verification scale interval (e) according
toDirective 2014/31/EU
mg 1 1 1 1
Minimum load (Min) according
toDirective 2014/31/EU
mg 0.01 0.1 0.1 0.1
Minimum initial weighing according
toUSP (United States Pharmacopeia),
Chap. 41
Optimum minimum initial weighing µg 82 820 820 820
Typical minimum initial weighing µg 300 820 820 820
Typical stabilization time s 7 5 5 5
Typical measurement time s 10 8 8 8
15.8 Recommended Calibration Weight
MCA2.7S MCA10.6S MCA6.6S MCA3.6P
Unit Value Value Value Value
External test weight g 2 10 5 3
Recommended accuracy class E2 E2 E2 E2
15.9 isoCAL Function
Unit Value
isoCAL is triggered
bythefollowingcriteria:
In the event of a temperature change K 1.5
After a time interval h 12
After successful leveling

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sartorius Cubis MCA2.7S and is the answer not in the manual?

Sartorius Cubis MCA2.7S Specifications

General IconGeneral
BrandSartorius
ModelCubis MCA2.7S
CategoryScales
LanguageEnglish

Related product manuals